Analog chlorine dioxide sensor CCS50

CCS50: Analog chlorine dioxide sensor ©Endress+Hauser
CCS50 analog chlorine dioxide sensor with adapter for installation in CCA250 flow assembly ©Endress+Hauser

Sensor for water, process water and utilities in all industries

CCS50 is a robust, low-maintenance sensor for chlorine dioxide measurement. It provides stable and fast measured values in water, process water and utilities. The sensor ensures efficient disinfection even at minimum water volumes and helps to achieve highest water quality and safety, to avoid overdosing of chlorine dioxide or to ensure the absence of ClO2 in beverage plants and reverse osmosis. CCS50 ensures continuity for the installed base of analog chlorine dioxide measuring points.

이 제품은 해당 국가에서 판매되지 않는 제품입니다.

빠른 선택

제안된 옵션 중에서 선택하십시오.
  • <%label%>
    • 더 많은 옵션을 찾고 계십니까? 계속하려면 개별 사양 구성 을 사용하십시오.
  • <%feature%>
      <%label%>
    • 더 많은 옵션을 찾고 계십니까? 계속하려면 개별 사양 구성 을 사용하십시오.
  • Process Connection
  • <%message%>

또는

개별 구성

가능한 모든 옵션에서 선택을 계속하십시오.
  • 장점

    • Fast response time (t90 < 15 s) provides an accurate process view and enables prompt reaction to process changes as well as efficient process control.

    • Increased process safety: precise and long-term stable measurement ensures consistent process...

  • 적용 분야

    CCS50 is an analog chlorine dioxide sensor for skid builders and end customers. It measures reliably in:

    • Drinking water - to ensure sufficient disinfection

    • Cooling systems or towers - to avoid pathogen growth

    • Wash water for packed vegetables and salads...

특징 및 사양

뒤로

Details

  • 측정 원리

    Chlorine dioxide

  • Application

    Guarantees reliable disinfection in drinking water
    Prevents the growth of biofilms and pathogens in cooling water
    Ensures food quality
    Ensures the absence or presence of chlorine dioxide in Utilities

  • Characteristic

    Amperometric measurement of dissolved chlorine dioxide

  • Measurement range

    Trace: 0 to 5 mg/l ClO2
    Standard: 0 to 20 mg/l ClO2

  • Measuring method

    Closed, membrane covered measuring cell
    Reduction of chlordioxide at the cathode

  • Design

    Closed amperometric 2-electrode measuring cell with PVDF membrane

  • Material

    Sensor shaft: PVC
    Membrane: PVDF
    Membrane cap: PVDF
    Sealing ring: FKM

  • Dimension

    Diameter: 25 mm (0.98 inch)
    Length: 126 mm (4.96 inch)

  • Process temperature

    0 to 55 °C (32 to 130°F), non-freezing

  • Process pressure

    Max.1 bar (max. 14.5 psi)

  • Temperature sensor

    10k NTC

  • Connection

    Analog with fixed cable