키워드, 주문 또는 제품 코드 또는 일련 번호를 검색해 보십시오(예: “CM442” 또는 “기술 정보”).
검색하려면 최소 2자 이상 입력하십시오.
Cerabar PMP71 - 절대압 및 게이지압 측정

절대압 및 게이지 압력계
Cerabar PMP71

메탈 센서가 장착된 압력 측정 계기

PMP71
Please Wait (spinning wheel)
가격 로드 중
일시적으로 가격을 로드할 수 없습니다.
Price on request
가격 활성화 중
가격: 정보 제공
from CHF ??.-
측정/작업 필터링
측정값

이 제품의 CAD 도면을 다운로드하도록 설정

  • 측정 원리

    Absolute and gauge pressure

  • Characteristic

    Digital transmitter with piezoresistive sensor and welded metallic membrane
    Modular transmitter
    Long term stability
    Enhanced safety via self diagnostic functionsSecondary process barrier

  • Supply voltage

    4...20 mA HART
    10,5...45V DC (Non Ex):
    Ex ia: 10,5...30V DC
    PROFIBUS PA:
    9...32 V DC (Non Ex)
    FOUNDATION Fieldbus:
    9...32 V DC (Non Ex)

  • Reference Accuracy

    Standard: 0.05%
    Platinum: up to 0.025%

  • Long term stability

    0.05 % of URL/ year
    0.07 % of URL/ 5 years
    0.1 % of URL/ 10 years

  • Process temperature

    -40°C...125°C
    (-40°F...257°F)

  • Ambient temperature

    -50°C...85°C
    (-58°F...185°F)

  • Measuring cell

    400 mbar...700 bar
    (5.8 psi...10.500 psi)
    relative/ absolute

  • Smallest calibratable span

    5 mbar (0.075 psi)

  • Vacuum resistance

    10 mbar (0.15 psi)

  • Max. Turn down

    100:1

  • Max. overpressure limit

    1050 bar (15.750 psi)

  • Process connection

    Thread:
    G1/2...G2, R1/2, MNPT1/2...MNPT2, R1/2
    Flange:
    DN25...DN80,
    ASME 1"...4",
    JIS 10K...20K

  • Material process membrane

    316L, AlloyC,
    Rhodium> Gold

  • Material gasket

    None, measuring cell welded

  • Fill fluid

    Silicone oil,
    Inert oil

  • Material housing

    Die-cast aluminum,
    AISI 316L

  • Communication

    4...20 mA HART
    1- 5V DC Low Power
    PROFIBUS PA
    FOUNDATION Fieldbus

  • Certificates / Approvals

    ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC, UK Ex

  • Safety approvals

    SIL

  • Design approvals

    EN10204-3.1
    NACE MR0175

  • Marine approvals

    GL/ABS

  • Drinking water approvals

    NSF

  • Specialities

    Diagnostic functions

  • Successor

    PMP71B

  • 측정 원리

    Absolute and gauge pressure

  • Characteristic / Application

    Digital transmitter with piezoresistive sensor and welded metallic membrane
    Modular transmitter
    Long term stability
    Enhanced safety via self diagnostic functions
    Secondary process barrier

  • Specialities

    diagnostic functionalities
    different languages in software

  • Supply / Communication

    4...20mA HART:
    10,5...45V DC
    Ex ia: 10,5...30V DC
    PROFIBUS PA /
    FOUNDATION Fieldbus:
    9...32V DC

  • Accuracy

    Standard: 0.05%
    Platinum: up to 0.025%

  • Long term stability

    0,05% of URL/year

  • Ambient temperature

    -50°C...85°C
    (-58°F...185°F)

  • Process temperature

    -40°C...125°C
    (-40°F...257°F)

  • Process pressure / max. overpressure limit

    1050bar (15,750psi)

  • Pressure measuring range

    400 mbar...700 bar
    (5.8 psi...10.500 psi)

  • Main wetted parts

    Alloy C276
    316L
    Rhodium

  • Process connection

    Threads
    Flanges (DIN, ASME, ...) with flush membrane

  • Max. measurement distance

    7000m (22.966ft) H2O

  • Communication

    4...20 mA HART
    1- 5V DC Low Power
    PROFIBUS PA
    FOUNDATION Fieldbus

  • Certificates / Approvals

    ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC

  • Safety approvals

    SIL

  • Design approvals

    EN 10204-3.1
    NACE MR0175, MR0103

  • Marine approval

    GL/ ABS

  • Drinking water approvals

    NSF

  • Options

    HistoROM/M-Dat
    4-line digital display
    SS- or Aluminiumhousing
    Separate housing

  • Successor

    PMP71B

  • Application limits

    Measuring cell: Metal
    welded
    If pressurized, possibly
    use differential pressure
    meas-urement with two
    pressure transmitters
    (electronic dp). Observe
    ratio head pressure :
    hydrostatic pressure